Skip to content

Characteristics of microSilicon diode detector for electron beam dosimetry

MetadataDetails
Publication Date2021-09-08
JournalJournal of Radiation Research
AuthorsYuichi Akino, Indra J. Das, Masateru Fujiwara, Akari Kaneko, Takashi Masutani
InstitutionsNorthwestern University, The University of Osaka
Citations5
AnalysisFull AI Review Included
  • Core Value Proposition: Comprehensive dosimetric characterization of the PTW microSilicon (type 60023) diode detector against the established Diode E and microDiamond for clinical electron beam radiotherapy.
  • Key Application: Electron beam commissioning, quality assurance, and profile measurements for energies ranging from 6 to 18 MeV on a Varian TrueBeam linac.
  • Critical Finding: microSilicon demonstrates minimal energy-dependent response, providing percent depth dose (PDD) curves without the stopping-power ratio correction required for ionization chambers.
  • Spatial Resolution: The 1.5 mm diameter active volume produces penumbra profiles nearly identical to the microDiamond and superior to the older Diode E.
  • Operational Stability: Negligible dose-rate dependence (<0.2%) across the tested range, a significant improvement over the logarithmic dependency observed in Diode E.
  • Cone Factor Accuracy: Cone factors measured with microSilicon are within ±1% of ionization chamber values across all clinical electron energies and cone sizes (5-25 cm).
  • Methodology: Direct comparison in a BluePhantom² water tank, validating PDD, off-center ratio (OCR), linearity, and dose-rate response per TRS-398 protocols.
ParameterValueUnitContext
Detector ModelmicroSilicon 60023—PTW unshielded p-type silicon diode
Sensitive Area1.5mm ØDisk-shaped active volume
Sensitive Thickness18µmThinner than Diode E (30 µm)
Entrance Window Density92mg/cm²Lower than Diode E (140 mg/cm²)
Water-Equivalent Window0.9mmRW3/Epoxy layered construction
EPOM (Electron)0.3mmDistance from detector tip
EPOM (Photon)0.9mmDistance from detector tip
Energies Tested6, 9, 12, 15, 18MeVClinical electron beam range
SSD100cmSource-to-surface distance
Applicator10 × 10cm²Reference collimation condition
Max Depth Variation (R50, R90, Rp)0.9mmCompared across all detectors/energies
Bremsstrahlung Dose Variation< 0.3%Inter-detector consistency
OCR Agreement< 1%Flat and tail regions
Dose-Rate Dependency< 0.2%microSilicon vs. log-dependence for Diode E
Cone Factor Deviation±1%Across 5-25 cm cones
Linearity Range1-1000MUMonitor unit response
  • Detector Selection: PTW microSilicon (60023), Diode E (60017), microDiamond (60019), and IBA PPC40 plane-parallel ionization chamber.
  • Beam Source: Varian TrueBeam linear accelerator at 6, 9, 12, 15, and 18 MeV.
  • Water Tank Setup: BluePhantom² scanning system with OmniPro software; SSD fixed at 100 cm; 10 × 10 cm² applicator.
  • EPOM Determination: Reference point set at zero; linear fitting applied to in-air data (−35 to −10 mm) and build-up region (3-9 mm); intersection calculated from three independent measurements.
  • PDD Measurement: PDI from PPC40 converted to PDD via TRS-398 water-air stopping power ratio correction; diode detectors used offset EPOM values.
  • Quantitative PDD Metrics: R50, R90, Rp, and Dx (X-ray contamination dose) extracted for all detectors.
  • OCR Measurement: Off-center ratios collected across the full field width for penumbra and flat-region analysis.
  • Linearity Test: Response evaluated from 1 to 1000 MU at constant dose rate.
  • Dose-Rate Test: Response measured from 0 to 1000 MU/min across all energies.
  • Cone Factor Test: Output factors measured for cone sizes 5, 10, 15, 20, and 25 cm.
  • Applicable Materials: The microDiamond detector used in this study relies on synthetic Single Crystal Diamond (SCD) as its active sensing element. For replication or extension of this dosimetry research, 6CCVD provides Optical Grade SCD with thicknesses from 0.1 µm to 500 µm, polished to Ra &lt; 1 nm, suitable for high-sensitivity radiation detection with minimal angular deviation. For applications requiring larger active areas, Polycrystalline Diamond (PCD) plates up to 125 mm diameter are available.

  • Customization Potential: The microDiamond detector specifies a 2.2 mm Ø, 1 µm thick sensitive volume with specific electrode geometry. 6CCVD supports custom metallization (Ti, Pt, Au, Pd, W, Cu) applied in-house to diamond substrates, enabling tailored Schottky or ohmic contact configurations. Precision laser cutting and dicing services ensure tight dimensional tolerances for detector fabrication.

  • Engineering Support: The study highlights diamond’s near-constant water-to-diamond stopping power ratio—a property directly tied to crystal quality. 6CCVD’s in-house PhD team can assist with material selection for radiation detection, beam monitoring, or spectroscopy projects, providing guidance on crystal orientation, doping strategies, and electrode optimization.

  • Call to Action: For custom specifications, metallization stacks, or material consultation on diamond-based radiation detector development, visit 6ccvd.com or contact our engineering team directly.

View Original Abstract

Abstract A microSilicon™ (PTW type 60023), a new unshielded diode detector succeeding Diode E (model 60017, PTW), was characterized for electron beam dosimetry and compared with other detectors. Electron beams generated from a TrueBeam linear accelerator were measured using the microSilicon, Diode E, and microDiamond synthetic single-crystal diamond detector. Positional accuracy of microSilicon was measured by data collected in air and water. The percent depth dose (PDD), off-center ratio (OCR), dose-response linearity, dose rate dependence, and cone factors were evaluated. The PDDs were compared with data measured using a PPC40 plane-parallel ionization chamber. The maximum variations of depth of 50% and 90% of the maximum dose, and practical depth among all detectors and energies were 0.9 mm. The maximum variations of the bremsstrahlung dose among all detectors and energies were within 0.3%. OCR showed good agreement within 1% for the flat and tail regions. The microSilicon detector showed a penumbra width similar to microDiamond, whereas Diode E showed the steepest penumbra shape. All detectors showed good dose-response linearity and stability against the dose rate; only Diode E demonstrated logarithmic dose rate dependency. The cone factor measured with microSilicon was within ±1% for all energies and cone sizes. We demonstrated that the characteristics of microSilicon is suitable for electron beam dosimetry. The microSilicon detector can be a good alternative for electron beam dosimetry in terms of providing an appropriate PDD curve without corrections, high spatial resolution for OCR measurements and cone factors.

  1. 2004 - Prescribing, recording, and reporting electron beam therapy (ICRU Report 71)
  2. 2009 - Recommendations for clinical electron beam dosimetry: supplement to the recommendations of Task Group 25 [Crossref]
  3. 2003 - The IPEM code of practice for electron dosimetry for radiotherapy beams of initial energy from 4 to 25 MeV based on an absorbed dose to water calibration [Crossref]
  4. 2006 - Intraoperative radiation therapy using mobile electron linear accelerators: report of AAPM Radiation Therapy Committee Task Group No. 72 [Crossref]
  5. 2019 - Breast intraoperative electron radiotherapy: Image-based setup verification and in-vivo dosimetry [Crossref]
  6. 1984 - Stopping powers for electrons and positrons (ICRU Report 37)
  7. 1985 - Characteristics of a p-Si detector in high energy electron fields [Crossref]
  8. 1987 - Practical methods of electron depth-dose measurement compared to use of the NACP design chamber in water [Crossref]
  9. 2001 - Determination of percentage depth-dose curves for electron beams using different types of detectors [Crossref]
  10. 2005 - Dosimetric characteristics of a new unshielded silicon diode and its application in clinical photon and electron beams [Crossref]