Skip to content

All-optical thermometry using a single multimode fiber endoscope and diamond nanoparticles containing nitrogen vacancy centers

MetadataDetails
Publication Date2022-08-01
JournalReview of Scientific Instruments
AuthorsLisa Ishikawa, T. Shikama, Takayuki Kakuno, Takato Watanabe, Mazin Jouda
InstitutionsKyoto University, Karlsruhe Institute of Technology
Citations3

Photoluminescence (PL) spectra from diamond nanoparticles containing negative nitrogen vacancy centers were measured by using a single multimode fiber endoscope combined with a high-sensitivity spectroscopy system. A laser light spot was produced at the distal end of the endoscope and the PL spectra from a temperature-controlled ensemble of diamond nanoparticles were measured. After calibrating the sensitivity and wavelength of the spectroscopy system, the temperature dependence of the zero-phonon line peak wavelength similar to those previously reported was obtained.