Infrared vertical external cavity surface emitting laser threshold magnetometer
At a Glance
Section titled “At a Glance”| Metadata | Details |
|---|---|
| Publication Date | 2024-02-26 |
| Journal | Applied Physics Letters |
| Authors | Nathan S. Gottesman, Michael A. Slocum, Gary A. Sevison, Michael Wolf, Michal L. Lukowski |
| Institutions | University of Arizona, Wright-Patterson Air Force Base |
| Citations | 6 |
| Analysis | Full AI Review Included |
Infrared VECSEL NV-Diamond Magnetometer: Technical Documentation
Section titled “Infrared VECSEL NV-Diamond Magnetometer: Technical Documentation”Executive Summary
Section titled “Executive Summary”- Core Innovation: The paper demonstrates the first infrared Laser Threshold Magnetometer (LTM) utilizing a vertical external cavity surface emitting laser (VECSEL) architecture integrated with an NV-center doped single crystal diamond (SCD) plate.
- Material Application: A 3 mm × 3 mm × 500 µm SCD plate with an NV density of 300 ppb is utilized as an intracavity saturable absorber at 1042 nm, enhancing the 1042 nm singlet-state absorption transition.
- Operating Wavelengths: The VECSEL operates at 1042.7 nm (single-frequency), pumped optically by an 808 nm pump laser, with NV polarization driven by a counter-propagating 532 nm pump.
- Key Performance Metrics:
- Best magnetic sensitivity: 7.5 nT/√Hz (10-50 Hz band, limited by VECSEL technical noise).
- Best projected PSNL sensitivity: 26.6 pT/√Hz.
- ODMR contrast: 18.4% at the transition frequency of 2.8229 GHz.
- Critical Observations: Near-threshold operation yields a nonlinear contrast enhancement attributed to an unexpected saturable absorption behavior, which improves both contrast and projected PSNL sensitivity beyond theoretical models.
- Engineering Bottleneck: Device performance is currently limited by VECSEL amplitude stability and technical noise rather than quantum spin projection noise.
Technical Specifications
Section titled “Technical Specifications”| Parameter | Value | Unit | Context / Condition |
|---|---|---|---|
| NV Density | 300 | ppb | Single crystal diamond intracavity plate |
| Spin Dephasing Time (T2*) | 1 | µs | Characteristic ensemble coherence time |
| Diamond Plate Dimensions | 3 × 3 × 0.5 | mm × mm × mm | Length × Width × Thickness |
| Zero-Field Splitting (D) | ≈ 2.87 | GHz | Ground state triplet splitting |
| ODMR Transition Frequency | 2.8229 | GHz | ∆ms = −1 transition (biased) |
| Bias Magnetic Field | 1.682 | mT | Applied via permanent ring magnets |
| Hyperfine Constant | 2.158 | MHz | 14N nuclear interaction |
| Resonance Linewidth (∆ν) | 1.98 ± 0.0508 | MHz | Power-independent Lorentzian width |
| Laser Wavelength | 1042.7 | nm | VECSEL single-frequency operation |
| 532 nm Pump Power | 1.2 | W | NV spin polarization source |
| 808 nm Pump Threshold (PThresh) | 4.33 | W | VECSEL pump power at threshold |
| TEM00 Mode Waist (2ws) | ≈ 80 | µm | Tight focus at diamond plane |
| Output Coupler Reflectivity (Rs) | 97 ± 0.5 | % | At 1042 nm |
| Cold Plate Temperature | 19.75 | °C | Maintained via Peltier cooler |
| Best Magnetic Sensitivity | 7.5 | nT/√Hz | 10 Hz to 50 Hz noise band |
| Best Projected PSNL Sensitivity | 26.6 | pT/√Hz | Limit imposed by photon shot noise |
| Maximum ODMR Contrast | 18.4 | % | Measured near lasing threshold |
| Equivalent Noise Bandwidth | ≈ 78 | Hz | At 1 ms LIA time constant |
| Output Optical Power (sensitivity test) | 28 | µW | Low-power technical noise characterization |
Key Methodologies
Section titled “Key Methodologies”- Diamond Sample Preparation:
- Fabricated a 3 mm × 3 mm × 500 µm SCD plate via chemical vapor deposition.
- Antireflection coating applied using a single layer of Al2O3 optimized for 1042 nm to suppress parasitic Fresnel reflections at the diamond facets.
- VECSEL Gain Chip Assembly:
- Processed a ½-VCSEL from an MOCVD-grown GaAs wafer.
- Structure: 12 compressively strained InGaAs quantum wells with GaAs barriers and strain-compensating GaAsP layers.
- Backed by 25 pairs of AlGaAs/AlAs DBR (Rs > 99.9%).
- DBR solder-bonded to a polycrystalline diamond heat spreader.
- Substrate removed via wet chemical etching; AR coating of single SiO2 layer applied via electron beam evaporation for 808 nm.
- Cavity Geometry (“V-Shaped”):
- Composed of the diamond sample, 1 mm BRF, 50 mm radius of curvature HR mirror, and a plane output coupler.
- Cavity tuned to 1042.7 nm using the BRF; single-frequency operation verified via 10 GHz scanning Fabry-Perot interferometer.
- Microwave loop antenna placed adjacent to the diamond surface.
- Optical Alignment & Detection:
- 532 nm pump delivered via dichroic mirror counter-propagating through the output coupler.
- 1042 nm output focused onto a silicon avalanche photodiode (250 mm focal length lens).
- 937 nm longpass filter placed before the detector to block NV photoluminescence.
- ODMR and Sensitivity Calibration:
- Permanent ring magnets biased the NV axis at 1.682 mT.
- Microwave probe (5 kHz square wave amplitude modulation, ≈ 25 dBm at the antenna) generated via SR386 signal generator and ZHL-5W-63S+ amplifier.
- Detection via SR865A lock-in amplifier combined with a SIM960 PID controller and ≈ 10 Hz RC filter for active VECSEL power stabilization.
- Magnetic sensitivity calculated from voltage-frequency conversion factors and averaged over 10 separate 10-second traces.
6CCVD Solutions & Capabilities
Section titled “6CCVD Solutions & Capabilities”The successful execution of an infrared VECSEL-based Laser Threshold Magnetometer depends critically on the optical, thermal, and quantum properties of the intracavity diamond plate. The following outlines how 6CCVD capabilities map directly to the requirements of this research.
-
Applicable Materials:
- Optical Grade SCD (Single Crystal Diamond): The paper utilized a 3 mm × 3 mm × 500 µm SCD plate. 6CCVD manufactures Optical Grade SCD with thicknesses from 0.1 µm to 500 µm, perfectly matching the intracavity geometry. Our bulk substrates (up to 10 mm) also support future iterations requiring thicker sensing volumes for higher NV interaction lengths.
- Custom NV Doping Profiles: While the paper used 300 ppb, 6CCVD works closely with research clients to provide tailored nitrogen-vacancy ensembles optimized for specific T2* targets and infrared absorption cross-sections.
-
Customization Potential:
- Antireflection Coatings: The paper relied on a single-layer Al2O3 coating for 1042 nm. 6CCVD offers custom dielectric coating design and deposition to achieve sub-1% reflectance at specific lasing wavelengths, which is critical for suppressing parasitic cavity losses.
- Custom Metallization and Laser Cutting: For experiments requiring direct integration of microwave antennas or electrodes onto the diamond surface, 6CCVD provides in-house metallization with Au, Pt, Pd, Ti, W, and Cu, paired with precision laser cutting and shaping services.
- Ultra-Low Surface Roughness: Surface imperfections interact heavily with the tightly focused TEM00 mode (2ws ≈ 80 µm). 6CCVD provides sub-nanometer polishing (Ra < 1 nm for SCD), directly reducing scattering losses in demanding VECSEL cavity geometries.
-
Engineering Support:
- 6CCVD’s in-house PhD engineering team can assist with material selection, NV center density optimization, and AR coating design for quantum sensing, VECSEL, and intracavity spectroscopy applications.
- We routinely support clients pushing the boundaries of cavity-enhanced magnetometry, providing the precision diamond optics necessary to advance photon shot noise limited sensitivity.
-
Call to Action:
- For custom specifications, NV-doped sample requests, or material consultation, visit 6ccvd.com or contact our engineering team directly.
View Original Abstract
Nitrogen-vacancy (NV) centers have considerable promise as high sensitivity magnetometers; however, they are commonly limited by inefficient collection and low contrasts. Laser threshold magnetometry (LTM) enables efficient collection and high contrasts, providing a path toward higher sensitivity magnetometry. We demonstrate an infrared LTM using an ensemble of NV centers in a single-crystal diamond plate integrated into a vertical external cavity surface emitting laser. The laser was tuned to the spin-dependent absorption line of the NV centers, allowing for optical readout by monitoring the laser output power. We demonstrate a magnetic sensitivity of 7.5 nT/Hz in the frequency range between 10 and 50 Hz. Furthermore, the contrast and the projected photon shot noise limited (PSNL) sensitivity are shown to improve significantly by operating close to the lasing threshold, achieving 18.4% and 26.6 pT/Hz near the threshold. In addition, an unexpected saturable absorption phenomenon was observed near the threshold, which enhanced the contrast and projected PSNL sensitivity.
Tech Support
Section titled “Tech Support”Original Source
Section titled “Original Source”References
Section titled “References”- 2017 - Magnetoencephalography for brain electrophysiology and imaging [Crossref]
- 2005 - The historical development of the magnetic method in exploration [Crossref]
- 2020 - Magnetic field fingerprinting of integrated-circuit activity with a quantum diamond microscope [Crossref]
- 2020 - Broadband multi-magnon relaxometry using a quantum spin sensor for high frequency ferromagnetic dynamics sensing [Crossref]
- 2016 - Absolute positioning using the Earth’s magnetic anomaly field: Magnetic anomaly [Crossref]
- 2020 - Sensitivity optimization for NV-diamond magnetometry [Crossref]
- 2020 - Diamond magnetometer enhanced by ferrite flux concentrators [Crossref]
- 2023 - High sensitivity of diamond nitrogen-vacancy magnetometer with magnetic flux concentrators via enhanced fluorescence collection [Crossref]
- 2023 - Nuclear quadrupole resonance spectroscopy with a femtotesla diamond magnetometer [Crossref]
- 2010 - Monolithic diamond optics for single photon detection [Crossref]