Subtractive gamma discrimination technique for neutron diamond detectors operating in mixed radiation fields
At a Glance
Section titled “At a Glance”| Metadata | Details |
|---|---|
| Publication Date | 2025-12-29 |
| Journal | Acta Polytechnica CTU Proceedings |
| Authors | Ondřej Novák, Ondřej Huml, M. Kropík, Ľubomír Sklenka |
| Institutions | Czech Technical University in Prague |
| Analysis | Full AI Review Included |
Executive Summary
Section titled “Executive Summary”- Detection Innovation: A novel subtractive gamma discrimination technique was developed to measure thermal neutron flux changes during rod drop transients at the VR-1 training reactor, replacing costly pulse shape discrimination methods.
- Cost-Effective Hardware: The system uses a dual-detector configuration (one with a B4C conversion layer, one without) to mathematically subtract gamma and fast neutron signals, eliminating the need for expensive high-sampling-rate digitizers.
- Diamond Detector Specifications: Artificial single crystal diamond plates (5 mm × 5 mm × 0.5 mm) with 110 Å aluminum electrodes and 1.2-1.4 µm B4C conversion layers containing 10B for thermal neutron capture via (n,α) reactions.
- Pulse Stretching Amplifier: Custom-developed THS3061D-based amplifier extending 4-6 ns pulses to ~50 ns, enabling use of lower-cost 125-250 Ms s-1 digitizers while maintaining >20,000 imp s-1 throughput.
- Experimental Validation: Tested with AmBe neutron source and cadmium filter, demonstrating ~23% count rate reduction for thermal neutron detectors (N1, N2) versus gamma-only detectors (G1, G2) when thermal neutrons were filtered.
- Reactor Application: Deployed for point-wise neutron flux distribution measurements during SCRAM (Safety Control Rod Activation Mechanism) transients at a zero-power training reactor.
- Key Achievement: Successful discrimination of thermal neutrons in mixed radiation fields using only pulse height analysis rather than complex shape recording, dramatically reducing electronics costs.
Technical Specifications
Section titled “Technical Specifications”| Parameter | Value | Unit | Context |
|---|---|---|---|
| Diamond Crystal Dimensions | 5 × 5 × 0.5 | mm | Single crystal artificial diamond (Element Six) |
| Crystal Type | sCVD (Single Crystal) | — | Active detection volume |
| Aluminum Electrode Thickness | 110 | Å (~10 nm) | Surface electrodes for charge collection |
| B4C Conversion Layer Thickness | 1.2-1.4 | µm | Contains 10B for thermal neutron capture |
| Detector Bias Voltage | -400 | V | Applied via CAEN N1470 HV source |
| Native Pulse Duration | <6 | ns | Diamond detector output pulse length |
| Conversion Reaction | 10B(n,α)7Li | — | Thermal neutron detection mechanism |
| Alpha Particle Range | <10 | µm | In diamond crystal |
| Thermal Neutron Fraction | 25-33 | % | Of total detected interactions |
| Pulse Shape Discrimination Limit | ~200 | counts s-1 | With CAEN N6743 digitizer |
| Subtractive Method Count Rate | >20,000 | imp s-1 | Fast-performance digitizer capability |
| Amplifier Gain | ~6 | dB | Signal shaping amplifier |
| Amplifier Pulse Extension Factor | ~6 | × | Extends pulse from ~6 ns to ~50 ns |
| THS3061D Slew Rate | 7,000 | V µs-1 | Operational amplifier specification |
| THS3061D Unity-Gain Bandwidth | 300 | MHz | Operational amplifier specification |
| CAEN N6743 Sampling Rate | 3 | Gs s-1 | High-end digitizer (reference) |
| Red Pitaya SIGNALlab Sampling Rate | 250 | Ms s-1 | Lower-cost digitizer option |
| Red Pitaya STEMlab Sampling Rate | 125 | Ms s-1 | Deployed digitizer (final system) |
| Coaxial Cable Type | MULTICOMP PRO MINI-1.13 | — | 1.13 mm diameter, 50 Ω |
| Cable Length (Detector to Amplifier) | 5 | m | Distance from reactor core/vessel |
| Measurement Standard Deviation | <3 | % | AmBe source measurement uncertainty |
| Cd Filter Count Reduction (N1) | 23 | % | Thermal neutron filtering verification |
| Cd Filter Count Reduction (N2) | 12 | % | Thermal neutron filtering verification |
| Neutron Source Distance | ~13 | cm | From AmBe to detector face |
Key Methodologies
Section titled “Key Methodologies”-
Diamond Crystal Selection: Artificial single crystal CVD diamond plates (5 mm × 5 mm × 0.5 mm) procured from Element Six; selected for consistent charge collection and well-defined pulse characteristics.
-
Electrode Deposition: 110 Å aluminum layers deposited on both crystal faces via evaporator, serving as electrical contacts for high-voltage bias and charge collection.
-
Conversion Layer Deposition: B4C layers (1.2 µm and 1.4 µm) deposited on one or both sides of selected crystals using evaporator with integrated crystal thickness monitor; deposition performed simultaneously for multiple crystals using custom holder design.
-
Electrical Connection: Aluminum electrodes connected via conductive epoxy and thin wires; design enabled compact multi-detector arrays within confined reactor instrumentation space.
-
Signal Chain Assembly: Detector → thin coaxial cable (1.13 mm diameter, shielded braid) → Cividec C2-HV preamplifier (with HV bias) → custom pulse-shaping amplifier → Red Pitaya STEMlab digitizer with FPGA-based pulse counting.
-
Pulse Stretching Amplifier Development: Circuit simulation in TINA-TI; iterative hardware prototyping with THS3061D current-feedback op-amp; optimization of feedback capacitance (C1 <3 pF limitation overcome with C8 addition enabling 15 pF); final PCB with large grounded surface, coaxial internal cabling, and shielded enclosure.
-
Discrimination Methodology: Dual-detector subtractive approach—one detector with B4C (thermal neutron sensitive) paired with one without (gamma/fast neutron only); count rate subtraction yields pure thermal neutron signal.
-
FPGA Pulse Counting: Negative pulse height threshold discrimination implemented in FPGA firmware; threshold set to separate detector signal pulses from electronic noise.
-
Calibration Protocol: AmBe neutron source in polyethylene moderator box with cadmium filter for thermal neutron absorption verification; per-detector calibration required due to manufacturing variations in conversion layer thickness.
-
Reactor Deployment: Multi-line detection system installed at VR-1 zero-power training reactor for rod drop transient measurements; FPGA recording enables continuous high-rate data acquisition during SCRAM events.
6CCVD Solutions & Capabilities
Section titled “6CCVD Solutions & Capabilities”Applicable Materials
Section titled “Applicable Materials”The research described requires high-purity Optical Grade Single Crystal Diamond (SCD) plates with stringent specifications for radiation detection. The 5 mm × 5 mm × 0.5 mm geometry and the need for uniform charge collection, well-defined pulse shape, and low noise make this application ideally suited to 6CCVD’s SCD offerings:
- Optical Grade SCD — Excellent for particle detection due to high charge carrier mobility and uniform response; recommended for thermal neutron detector active volumes.
- Detector-Grade SCD Plates — Available in custom thicknesses (0.1 µm-500 µm range covers the 0.5 mm requirement with ample margin for thinner detection volumes or thicker absorber designs).
- Boron-Doped Diamond (BDD) — Could serve as an alternative active material for applications requiring intrinsic gamma/neutron sensitivity differentiation through dopant-controlled electrical properties.
Customization Potential
Section titled “Customization Potential”The paper’s detector design relies on several custom features that align directly with 6CCVD’s in-house capabilities:
- Custom Metalization: The 110 Å aluminum electrodes required for the diamond detectors can be replicated using 6CCVD’s internal metallization capabilities. Alternative metallization schemes (Ti/Pt/Au for wire bonding, or W for high-temperature applications) can be provided to match specific signal extraction or integration requirements.
- Laser Cutting Services: Custom diamond geometries beyond standard wafer formats—including the 5 mm × 5 mm squares used in this research—can be produced via precision laser cutting. This enables detector arrays, segmented detectors, or application-specific form factors.
- Surface Engineering: Pre-polished surfaces (Ra < 1 nm for SCD) ensure consistent electrode deposition and minimize surface charge trapping effects that degrade detector performance.
- Scalable Dimensions: While this study used 5 × 5 mm crystals, 6CCVD supports plates up to 125 mm (polycrystalline) and custom SCD sizes, enabling both single-point detectors and larger-area detection arrays for flux mapping applications.
Engineering Support
Section titled “Engineering Support”6CCVD’s in-house PhD engineering team can assist researchers developing CVD diamond-based radiation detection systems with:
- Material Selection Guidance: Matching diamond grade (SCD vs. PCD), purity level, and dopant concentration to specific radiation detection requirements—whether for thermal neutron detection with conversion layers, direct fast neutron sensing, or gamma spectroscopy.
- Electrode Design Consultation: Recommendations for electrode materials (Al, Ti, Pt, Au, W) and thicknesses optimized for charge collection, adhesion to diamond surfaces, and compatibility with downstream wire bonding or PCB integration.
- Geometry Optimization: Trade-off analysis between active volume thickness (affecting detection efficiency and pulse amplitude), lateral dimensions (affecting spatial resolution and count rate capability), and manufacturing yield.
- Conversion Layer Integration Advisory: While B4C deposition is typically performed by specialized facilities (as done at CEITEC in this study), 6CCVD can advise on substrate preparation, surface termination, and interface engineering to optimize conversion layer adhesion and uniformity.
Call to Action
Section titled “Call to Action”For custom specifications, material consultation, or to discuss diamond detector fabrication for nuclear instrumentation, reactor monitoring, or mixed-field radiation detection applications, visit 6ccvd.com or contact our engineering team directly.
View Original Abstract
A detection system based on diamond detectors and subtractive discrimination was developed for the measurement of thermal neutrons flux distribution changes during rod drop transient at the VR-1 reactor. The detection system was modified to replace the pulse-shaped method used for discrimination. The electronics available at the department was not able to measure high count rates with the pulse shape discrimination method. Subtractive discrimination allows us to reduce the requirements in the signal processing area, as the shape does not need to be recorded, and therefore the digitizer performance can be significantly lower. For this reason, a pulse stretching amplifier was developed and tested.