Application of boron-doped diamond modified cobalt oxide nanoparticles (BDD/CoONPs) as an electrochemical sensor for hydrocortisone detection
Paper details
Section titled “Paper details”| Field | Value |
|---|---|
| Journal | International Journal of Electrochemical Science |
| Published | 2026-02-20 |
| Authors | Yunita Triana, Irfan Riadin, Sarah Toding, Andi Idhil Ismail, Kartika A. Madurani |
| DOI | 10.1016/j.ijoes.2026.101329 |
A boron-doped diamond electrode modified with cobalt oxide nanoparticles detected hydrocortisone with a reported limit of detection of 8.19 × 10⁻³ μM. Differential pulse voltammetry in 0.1 M NaOH produced an oxidation peak at +0.75 V versus Ag/AgCl.
Performance against glassy carbon
Section titled “Performance against glassy carbon”The comparison used BDD, BDD/CoONPs, GC, and GC/CoONPs at 0.04 μM hydrocortisone in 0.1 M NaOH. GC/CoONPs served as the paper’s material reference for a hydrocortisone sensor built on glassy carbon. Unmodified BDD and GC produced no detectable hydrocortisone peaks. The modified electrodes produced oxidation signals, with the diamond-based electrode giving the stronger response.
| Measure | BDD/CoONPs | GC/CoONPs | Test condition |
|---|---|---|---|
| Oxidation current density | About 6.5 μA cm⁻² | About 0.95 μA cm⁻² | 0.04 μM HC, 0.1 M NaOH |
| Calibration linearity | R² = 0.99 | R² = 0.98 | 0.04 to 0.20 μM HC |
| Limit of detection | 8.19 × 10⁻³ μM | 2.39 × 10⁻² μM | 0.1 M NaOH calibration |
| Repeatability | 1.43% RSD | 3.35% RSD | 0.04 μM HC, pH 12, ten trials |
| Reproducibility | 1.03% RSD | 6.87% RSD | 0.003 μM HC, ten days at 25 °C |
The diamond electrode’s stronger response was consistent with BDD’s reported low background current, chemical inertness, mechanical robustness, and resistance to surface fouling. The authors also linked uniform CoONPs distribution on BDD to sharper and more reproducible oxidation peaks. That distribution was proposed from the voltammograms rather than measured directly.
Electrode preparation and measurement conditions
Section titled “Electrode preparation and measurement conditions”Before modification, BDD was immersed in aqua regia with a 3:1 HCl:HNO₃ ratio for 30 min. The electrode was then sonicated in ethanol for 15 min and dried with nitrogen.
A 5% Nafion solution was diluted with ethanol to 0.5%. Two mg CoONPs were mixed with the diluted Nafion and sonicated for 1 h. The mixture was dripped onto BDD and dried at room temperature.
Measurements used an Autolab PGSTAT101 in a three-electrode cell. The working electrode was BDD/CoONPs, the counter electrode was BDD, and the reference was Ag/AgCl with 3 M KCl. The BDD working area was 0.269 cm². Each measurement used 10 mL of working solution. Hydrocortisone was 0.04 μM unless another concentration is stated. DPV ran from +0.5 to +1.6 V versus Ag/AgCl.
CoONPs formed below 5 nm
Section titled “CoONPs formed below 5 nm”CoONPs were produced with Nigella sativa seed extract as the reducing and stabilizing agent. Five grams of mashed seeds were heated in 100 mL distilled water at 80 °C for 15 min. Cobalt nitrate at 0.02 M was added, and the mixture was stirred at 80 °C and 600 rpm for 30 min. After standing at room temperature for 24 h, the precipitate was filtered and dried at 80 °C for 10 h.
Transmission electron microscopy showed particles under 5 nm, with a mean size of 1.72 nm and standard deviation of 0.55 nm. XRF showed a cobalt peak above 30,000 counts at approximately 6.9 keV.
UV-Vis absorption peaks at 320 and 344 nm were assigned to charge transfer between Co²⁺ and organic seed ligands. FTIR bands at 2921 and 2851 cm⁻¹ indicated C-H stretching from alcohol or organic ligands. Bands at 700 and 663 cm⁻¹ were assigned to Co-O vibrations.
These data supported cobalt-containing oxide nanoparticles with organic seed ligands. The reported characterization included TEM, FTIR, XRF, and UV-Vis, but no diffraction-based phase identification.
Oxidation current peaked at pH 11 and 12
Section titled “Oxidation current peaked at pH 11 and 12”Hydrocortisone oxidation was tested at 0.04 μM across pH 8, 9, 10, 11, and 12. The highest peak currents occurred at pH 11 and 12. Most sensor tests used the alkaline condition of 0.1 M NaOH at pH 12.
Peak-potential regressions gave slopes of -0.307 × 10⁻⁶ V/pH for BDD/CoONPs and -0.365 × 10⁻⁶ V/pH for GC/CoONPs. The corresponding R² values were 0.9917 and 0.9483.
The authors linked the pH-dependent shifts to the Nernst relation. They also stated that higher H⁺ activity reduced hydrocortisone oxidation at lower pH.
The authors propose a Co²⁺/Co³⁺ oxidation pathway
Section titled “The authors propose a Co²⁺/Co³⁺ oxidation pathway”The authors present the oxidation route as a plausible hypothesis, not a directly measured pathway. Under applied potential, hydrocortisone is proposed to oxidize at a hydroxyl group on its steroid ring. The product is described as a more reactive radical or oxidized species that releases one electron. The paper represents this first step as hydrocortisone losing an electron to form HC⁺.
Co²⁺ is proposed to oxidize to Co³⁺ during electron transfer, while Co³⁺ can accept an electron and return to Co²⁺. CoONPs may promote hydrocortisone adsorption through interactions involving hydroxyl and carbonyl groups with cobalt ions.
The paper reports no direct measurement of the radical or operating cobalt valence during oxidation. The Co²⁺/Co³⁺ cycle therefore remains a proposed explanation for the observed current.
Calibration, precision, and interference
Section titled “Calibration, precision, and interference”Calibration used five hydrocortisone concentrations of 0.04, 0.08, 0.12, 0.16, and 0.20 μM. The BDD/CoONPs regression was y = 4.029E-05x + 4.950E-06. The GC/CoONPs regression was y = 2.651E-06x + 1.003E-06. The paper did not specify units for the regression coefficients.
Peak current increased with concentration, while peak potential remained nearly constant. The authors attributed rising background current to adsorption-induced surface charging and increased interfacial capacitance. They described this baseline shift as characteristic of adsorption-controlled electrochemical sensors.
Repeatability used one 0.04 μM solution over ten trials. Reproducibility was measured over ten consecutive days at 0.003 μM and 25 °C, with three repetitions.
The interference test combined 10 μM HC, 10 μM ascorbic acid, and 10 μM glucose in 0.1 M NaOH. The combined trace showed an ascorbic acid peak at +0.55 V, another peak at +0.62 V, and an HC peak at +0.75 V. All three analytes were present at 10 μM, so the test did not establish tolerance when interferents exceed HC.
Biological validation remains unresolved
Section titled “Biological validation remains unresolved”The sensor has not been validated in plasma or another biological matrix. The authors identify protein binding to transcortin and albumin as a factor that may reduce electrochemical availability. They also identify cortisone and progesterone as endogenous steroids that may overlap the HC signal. The authors state that plasma studies are needed to quantify matrix effects and clarify the electrochemical response.
The conclusion states that the LoD is below the physiological HC range, but gives no numeric range. The reported LoD also lies below the lowest calibration point of 0.04 μM, so validation near the limit was not shown.
The reproducibility section reports 0.003 μM HC, but the methods overview lists 0.04 μM for sensor measurements. The paper does not reconcile that concentration difference. The GC/CoONPs peak is reported at -0.38 V, outside the stated DPV scan of +0.5 to +1.6 V. The paper does not resolve this reporting inconsistency.
The study also omits the boron concentration, diamond grade, CoONPs mass loading per area, and coating thickness. Those omissions limit reproduction of the modified electrode. For material selection, the results support BDD as a stable electrode base for alkaline oxidation with CoONPs. Engineers should treat the LoD as a buffer result until near-limit calibration and plasma validation are complete.
Materials and support
Section titled “Materials and support”The sensor requires boron-doped diamond for its working and counter electrodes, including a surface that accepts the Nafion/CoONPs coating after chemical cleaning. 6CCVD offers MPCVD-grown single-crystal and polycrystalline boron-doped diamond through the diamond materials catalog, matching the paper’s electrode material for electrochemical sensing. The custom fabrication and engineering support service can help define electrode preparation and coating requirements for a sensor build.
View Original Abstract
In this study, a boron-doped diamond (BDD) electrode modified with cobalt oxide nanoparticles (CoONPs) is proposed as a working electrode for hydrocortisone (HC) detection in 0.1 M NaOH. CoONPs were synthesized via a green synthesis approach using Nigella sativa seed extract and cobalt nitrate as precursors. The synthesized CoONPs were characterized in terms of functional groups, elemental composition, and surface morphology. The electrochemical redox behavior of HC was investigated using differential pulse voltammetry (DPV) within a potential range of +0.5 to +1.6 V (vs. Ag/AgCl), with variations in pH and analyte concentration to evaluate the sensing performance and determine the limit of detection (LoD). A detection limit of 8.19 × 10⁻³ μM was achieved, accompanied by good measurement stability, with repeatability and reproducibility values of 1.43% and 1.03% RSD, respectively. These results demonstrate the strong potential of the BDD/CoONPs electrode for sensitive and reliable HC detection.
Boron Doped Diamond On Titanium Bdd Electrode
View Specs →
Boron Doped Diamond Electrode On Niobium Bdd Nb Electrode
View Specs →
Boron Doped Single Crystal Diamond Plate
View Specs →References we cover
Section titled “References we cover”Original source
Section titled “Original source”References
Section titled “References”- 2020 - Sweat and saliva cortisol response to stress and nutrition factors [Crossref]
- 2014 - Plasma, salivary and urinary cortisol levels following physiological and stress doses of hydrocortisone in normal volunteers [Crossref]
- 2024 - Salivary cortisol measurement contamination after oral hydrocortisone: a randomized crossover trial [Crossref]
- 2019 - Development and validation of analytical spectrophotometric and RP-HPLC methods for the simultaneous estimation of hydroquinone, hydrocortisone and tretinoin ternary mixture in topical formulation [Crossref]
- 2017 - Paper-based analytical biosensor chip designed from graphene-nanoplatelet-amphiphilic-diblock-co-polymer composite for cortisol detection in human saliva [Crossref]
- 2021 - Development of a highly sensitive imprinted electrochemical sensor for the detection of hydrocortisone in wastewater [Crossref]
- 2015 - A new approach for quantitative determination of γ-cyclodextrin in aqueous solutions: application in aggregate determinations and solubility in hydrocortisone/γ-cyclodextrin inclusion complex [Crossref]
- 2018 - Comparative study of two different chromatographic approaches for quantitation of hydrocortisone acetate and pramoxine hydrochloride in presence of their impurities [Crossref]
- 2020 - Novel eco-friendly chromatographic determinations of hydrocortisone acetate, fusidic acid, their pharmacologically active impurities and pharmaceutical excipients: a comparative study [Crossref]
- 2024 - Current perspective of advanced electrochemical oxidation processes in wastewater treatment and life cycle analysis [Crossref]