FieldValue
JournalElectrophoresis
Published2026-05-02
AuthorsPetr Kubáň, Jiří Volánek, Nguyen Thi Thu Trang, Тран Дай Лам
DOI10.1002/elps.70099

Reviewed CE-AD studies report a dopamine mass limit of about 400 ± 100 zmol in standard solution. Urine assays reach 0.3 to 0.6 nM for monoamines. The review links these gains to capillary miniaturization, electrode geometry, and online preconcentration. Across 77 surveyed articles, carbon electrodes account for more than 70% of reported detectors.

The authors searched Web of Science in October 2025 and selected 77 articles on CE-AD of neurotransmitters. They compared electrode materials, decoupling, capillary formats, background electrolytes, analytical limits, and sample matrices. The paper is a literature synthesis, not an interlaboratory study. Performance values therefore belong to their source methods.

CE separates analytes by electrophoretic mobility in a high electric field. Fused-silica capillaries usually have a 25 to 75 µm inner diameter and lengths of several tens of centimeters. Separation voltages reach 30 kV, with theoretical plate counts often reaching hundreds of thousands.

Uncoated silanol groups generate electroosmotic flow toward the detector. Buffer pH, ionic strength, and surface chemistry alter migration order and analysis time. Sample injections occupy 1% to 2% of capillary volume, usually at nanoliter scale.

At a fixed working-electrode potential, analytes oxidize or reduce and produce faradaic current proportional to concentration. Small electrodes lower capacitive charging noise, so miniaturization can preserve or improve signal-to-noise.

Catecholamines undergo two-electron oxidation of their ortho-dihydroxybenzene group to a quinone. Serotonin has an oxidizable 5-hydroxybenzene group. Histamine needs at least +1.1 V versus Ag/AgCl or a modified catalytic surface.

Separation voltages of 10 to 30 kV generate background currents from units to tens of microamperes. Target faradaic currents are often picoamperes. Poor isolation raises noise and can shift the working-electrode potential.

Off-column detection isolates the circuit with porous glass, Nafion, cellulose acetate, or etched junctions. It lowers noise, but junctions can cause band broadening, analyte loss, and mechanical complexity.

End-column detection uses the high resistance of narrow capillaries to suppress the field at the outlet. Inner diameters below 25 µm are common. Diameters below 10 µm and sub-micrometer formats have also been reported. A 50 to 75 µm capillary requires low conductivity and generally gives higher limits.

Wall-jet detection is robust and easy to align. Larger electrodes intercept more analyte, but increase detection volume and dispersion. A comparison of 25 and 500 µm disks found similar performance.

In-capillary placement removes postcolumn dead volume, but raises risks of breakage, bubbles, and trapped impurities. Picoampere detection also needs very low input-bias-current amplifiers.

Carbon fibers are typically 5 to 10 µm in diameter. The original Wallingford and Ewing system used a 10 µm fiber, a porous-glass decoupler, and CAPS at pH 10 in a 75 µm capillary. It produced 180,000 theoretical plates and picomole sensitivity.

Detector and methodSeparation or buffer conditionReported limitMain result
Carbon fiber, end-column9 µm inner diameter0.7 amol for serotoninHigh mass sensitivity, but narrow capillaries can clog
Pt disk, wall-jet5 µm capillary; 20 mM MES, pH 6.03.0 amol DA; 5.2 amol catecholPeak-current RSD was 2% to 5%
BDD microline, end-column30 mM MES, pH 5.7Below 25 nM for DA, NE, and E155,000 plates for DA and 0.5 to 1 pA noise
Carbon disk, dynamic pH junction150 mM boric acid, pH 10.335.34 to 68.3 nMAbout 100-fold signal gain in urine
Carbon fiber array, FASS160 mM phosphate, pH 5.90.3 to 0.6 nM in urineAbout 50-fold gain in urine

Mass limit, concentration limit, and fouling resistance measure different properties. BDD performed well in repeated tissue injections, and several studies required no polishing or electrochemical pretreatment. Its wide potential window and low background help when organic products would foul carbon. The review also notes higher cost, complex fabrication, and lower intrinsic sensitivity than carbon.

Modified carbon can lower operating potential and suppress selected interferents. Pt/MWCNTs@Pdop reached 0.3 to 0.92 nM for 5-HT, DA, NE, and E in rat hippocampus and hypothalamus. Modified layers can degrade, leach, and reduce reproducibility.

Redox cycling repeatedly oxidizes and reduces an analyte between generator and collector electrodes. A urine assay gave limits of 0.14 nM for E, 0.16 nM for NE, and 0.41 nM for DA. It used fivefold-diluted urine, 150 mM MES with sodium acetate at pH 5.57, and a 12 min separation. Dual electrodes add signal and selectivity.

Au supports microfabrication and surface chemistry, but has a narrower aqueous potential window and generally higher limits than carbon systems.

Phosphate is the most common background electrolyte. Studies use 10 to 200 mM phosphate at pH 6 to 7.4 because it buffers near physiological pH. Its ionic strength raises separation current and Joule heating.

MES is used at 20 to 30 mM and pH 5.5 to 6.5 for high-sensitivity catecholamine work. The review links MES to the lowest mass limits with carbon fiber microelectrodes.

Borate supports pH 8 to 10 separations and complexes catechol diols. It also supports SDS-based MEKC, although high-pH operation increases adsorption and fouling concerns.

FASS, dynamic pH junctions, and transient isotachophoresis increase concentration sensitivity. Weng’s FASS method gained about 50-fold in urine and 5,000-fold in standard solution. The dynamic pH method gained about 100-fold. These procedures also require dilution or cleanup to control urine salts and matrix effects.

A dual-electrode urine assay separated DA, E, and NE within 12 min. It quantified them in fivefold-diluted urine at limits of 0.14 to 0.41 nM. The FASS method reached 0.3 to 0.6 nM without offline enrichment.

A lymphocyte study compared 58 MS patients with 19 controls using 25 mM MES at pH 5.65. LODs were 0.13 fmol/mg DA, 0.37 fmol/mg NE, and 0.11 fmol/mg DOPAC. It measured elevated epinephrine in first-attack MS and reduced norepinephrine in relapsing-remitting MS.

Rat brain microdialysis with an integrated etched decoupler reached near 1 nM DA and used 1 min collections. It tracked extracellular DA after uptake inhibition and electrical stimulation. BDD rat-heart analysis used 250 mM borate at pH 8.8 and kept noise at or below 0.65 pA. It also measured regional NE differences.

Single-cell studies found 0.29 to 1.28 fmol DA per PC12 cell. The mean was 0.61 ± 0.30 fmol across 10 cells. A single Drosophila larval brain required about 8 nL homogenate. LODs were 1 nM for DA and 5-HT, 2.5 nM for TA, and 4 nM for OA.

Electrode fouling remains the main stability problem. Oxidative products can form polymeric or adsorbing species that passivate the surface. Signal falls and reproducibility worsens. Pulsed cleaning waveforms, Nafion coatings, melanin films, and fouling-resistant BDD help. Stable long-term operation in complex biological matrices remains unresolved.

High-voltage isolation imposes a second tradeoff. Physical decouplers lower noise but add interfaces. End-column devices remove the interface, but narrow capillaries and low-conductivity electrolytes reduce robustness or sensitivity.

Extreme nanocapillaries enable single-cell analysis but clog easily. Clinical translation also remains limited. Most evidence comes from animal models, invertebrate tissues, or a few human urine and lymphocyte studies. The authors propose disposable electrode-capillary modules, microfluidic integration, and wearable sampling for future work. These are proposed directions, not demonstrated clinical performance.

Material choice should follow the limiting requirement. Carbon fiber suits the lowest mass limits. Carbon disk suits a robust, polishable detector. BDD suits fouling-prone matrices and high-potential oxidation. Modified carbon suits lower-potential selectivity. Generator-collector arrays suit subnanomolar concentration targets. The concrete design task is to match capillary bore, buffer conductivity, decoupling, and fouling control to the required mass and concentration limits.

For the boron-doped diamond working electrodes reviewed here, 6CCVD’s diamond materials catalog lists MPCVD-grown boron-doped diamond and Pt metallization. These materials are relevant to the paper’s polycrystalline BDD-on-Pt detector concept. The CE-AD application needs a conductive, low-background electrode that resists fouling in biological matrices, and the linked engineering support can help define the film and Pt interface.

View Original Abstract

Capillary electrophoresis with amperometric detection has become a powerful tool for the determination of electroactive neurotransmitters due to its high separation efficiency, minimal sample consumption, and excellent detection sensitivity. This review provides the first comprehensive summary specifically on monoamine neurotransmitters and related compounds in biological systems. Fundamental principles of amperometric detection, electrode positioning strategies, and approaches to high-voltage decoupling are described, followed by a critical comparison of electrode materials including carbon fiber, glassy carbon, graphite, noble metals, boron-doped diamond, and chemically modified or multielectrode designs. Their impact on noise, sensitivity, fouling resistance, and ease of integration is highlighted. Particular attention is given to background electrolyte composition, where phosphate buffers are most widely used, while 2-(N-morpholino)ethanesulfonic acid systems have enabled the lowest limits of detection, often reaching the attomole range. Applications span human urine, blood, lymphocytes, brain tissue, microdialysates, invertebrate tissues, and single cells, demonstrating capillary electrophoresis with amperometric detection as a versatile platform for neurochemical analysis. Advances in online preconcentration, dynamic pH junctions, and redox-cycling electrodes have further improved selectivity and sensitivity. Finally, current limitations, such as electrode fouling and interface robustness, are discussed together with emerging trends, including clinical translation and in vivo neuroanalysis.