Skip to content

Bidimensional polycrystalline CVD diamond detector for Intensity Modulated Radiation Therapy pre-treatment verifications

MetadataDetails
Publication Date2015-03-25
JournalJournal of Instrumentation
AuthorsM. Zani, M. Scaringella, C. Talamonti, A. De Sio, E. Pace
InstitutionsUniversity of Florence, Istituto Nazionale di Fisica Nucleare, Sezione di Firenze
Citations4

This study aims at investigating the possible employment for pre-treatment verifications in Intensity Modulated Radiation Therapy (IMRT) of a polycrystalline Chemical Vapour Deposited (pCVD) diamond bidimensional detector. The pCVD device, with an area of 2.5 Ɨ2.5cm2 (12Ɨ 12 pixels), has been used to measure dose maps of a 10 MVRX prostatic IMRT field. Its response was compared both with a commercial bi-dimensional detector made with silicon and with Treatment Planning System (TPS) calculations. Measurement provided promising results on a map of 1.8Ɨ12.6cm2. Absorbed doses measured along IMRT profiles by our device are consistent with the ones acquired with the commercial device and an overall good agreement with respect to the TPS was found for the diamond dosimeter.