Bidimensional polycrystalline CVD diamond detector for Intensity Modulated Radiation Therapy pre-treatment verifications
At a Glance
Section titled āAt a Glanceā| Metadata | Details |
|---|---|
| Publication Date | 2015-03-25 |
| Journal | Journal of Instrumentation |
| Authors | M. Zani, M. Scaringella, C. Talamonti, A. De Sio, E. Pace |
| Institutions | University of Florence, Istituto Nazionale di Fisica Nucleare, Sezione di Firenze |
| Citations | 4 |
Abstract
Section titled āAbstractāThis study aims at investigating the possible employment for pre-treatment verifications in Intensity Modulated Radiation Therapy (IMRT) of a polycrystalline Chemical Vapour Deposited (pCVD) diamond bidimensional detector. The pCVD device, with an area of 2.5 Ć2.5cm2 (12Ć 12 pixels), has been used to measure dose maps of a 10 MVRX prostatic IMRT field. Its response was compared both with a commercial bi-dimensional detector made with silicon and with Treatment Planning System (TPS) calculations. Measurement provided promising results on a map of 1.8Ć12.6cm2. Absorbed doses measured along IMRT profiles by our device are consistent with the ones acquired with the commercial device and an overall good agreement with respect to the TPS was found for the diamond dosimeter.