Skip to content

Investigation of the diamond based detectors characteristics with different thickness of the sensor element

MetadataDetails
Publication Date2017-01-01
JournalJournal of Physics Conference Series
AuthorsE. V. Gladchenkov, R. F. Ibragimov, V.A. Kolyubin, P. G. Nedosekin, E. M. Tyurin
InstitutionsNational Research Nuclear University MEPhI

This work is devoted to study of the diamond based radiation detectors. Experiments were carried out with two types of detectors: based on a thin diamond film and on a composite diamond plate. The following types of ionizing radiation has been used in experiments: beta radiation of 90Sr - 90Y, fission fragments and alpha particles of 252Cf, and Kr ions obtained at the particle accelerator. It is shown that the developed thin-film diamond based detector effectively registers heavy charged particles, whereas beta, neutron and gamma radiation does not give a significant contribution to the detector signals. Those type of detectors are proposed for measurement of heavy charged particles linear energy transfer in diamond. The multilayer diamond based detector (detector with a composite diamond plate) showed improved charge collection efficiency values in compare with the detection with a single diamond plate.