Skip to content

Arsenic and arsenate detection with anodic stripping voltammetry technique using boron-doped diamond modified by platinum

MetadataDetails
Publication Date2020-02-01
JournalIOP Conference Series Materials Science and Engineering
AuthorsRinaldo Sitorus, Asep Saefumillah, Tribidasari A. Ivandini
InstitutionsUniversity of Indonesia
Citations2
AnalysisFull AI Review Included
  • A Platinum-modified Boron-Doped Diamond (Pt-BDD) electrode was successfully fabricated via a two-step process combining wet chemical seeding and amperometric electrodeposition, achieving a Pt surface coverage of 1.54% as confirmed by SEM-EDS.
  • The modified electrode was applied to electrochemical detection of arsenic species (As3+ and As5+) using Anodic Stripping Voltammetry (ASV) in a 0.1 M phosphate buffer solution (pH 6).
  • Optimized detection parameters include a deposition potential of -500 mV, a deposition time of 150 s, and a scan rate of 200 mV/s.
  • A critical pretreatment step using 0.1 M NaBH4 was employed to chemically reduce As5+ to As3+ prior to measurement, enabling speciation analysis of mixed arsenic solutions.
  • The sensor demonstrated strong linearity (R2 = 0.9797 for As3+; R2 = 0.9903 for As5+) across the 0-100 ppb range, with estimated limits of detection of 16.50 ppb for As3+ and 8.19 ppb for As5+.
  • Excellent operational stability was verified, with a relative standard deviation (%RSD) below 3% over five consecutive days of testing, confirming the electrode’s suitability for field-deployable water quality monitoring.
ParameterValueUnitContext
BDD SubstrateB-doped diamond—Working electrode base
Pt Coverage1.54%Atomic surface coverage (SEM-EDS)
Modification MethodWet seeding + Electrodeposition—Two-step Pt deposition
Electrodeposition Potential-0.2VPt growth phase
Electrodeposition Time15minDuration of Pt growth
Calcination Temperature700°CPost-deposition annealing
Calcination AtmosphereN2—Inert environment, 5 min duration
Buffer SystemPhosphate Buffer Solution (PBS)—Supporting electrolyte
Buffer Concentration0.1MpH 6
ReductantNaBH4—0.1 M, converts As5+ → As3+
Stirring Time (Reduction)1minPost-NaBH4 addition
Deposition Potential (ASV)-500mVPreconcentration step
Deposition Time (ASV)150sPreconcentration step
Scan Rate (ASV)200mV/sStripping step
Linear Range (As3+)0-100ppbCalibration range
Linear Range (As5+)40-100ppbCalibration range
R2 (As3+)0.9797—Linearity coefficient
R2 (As5+)0.9903—Linearity coefficient
LOD (As3+)16.50ppbEstimated limit of detection
LOD (As5+)8.19ppbEstimated limit of detection
Stability (RSD)< 3%Over 5 consecutive days
  • BDD Surface Preparation: B-doped diamond substrates supplied (Einaga’s Lab, Keio University); surface was the starting template for Pt modification.
  • Wet Chemical Seeding: Sequential drop-casting of 0.1 M NaBH4 and H2PtCl6 solutions onto the BDD surface; rested for 24 hours, then washed and dried. Repeated three times.
  • Electrodeposition: Performed at -0.2 V for 15 minutes in H2PtCl6 solution to grow Pt crystal nuclei into larger, homogeneously distributed particles.
  • Thermal Annealing: Calcination at 700 °C under N2 atmosphere for 5 minutes to stabilize Pt particles.
  • Surface Characterization: SEM imaging at 2000× and 10,000× magnification; SEM-EDS for atomic coverage analysis.
  • ASV Detection Protocol (As3+): Direct anodic stripping voltammetry in PBS pH 6; deposition at -500 mV for 150 s; stripping at 200 mV/s.
  • ASV Detection Protocol (As5+): Addition of 0.1 M NaBH4, stirred for 1 minute to reduce As5+ → As3+; followed by identical ASV parameters.
  • Mixed Species Detection: Two-step sequential ASV—first measures As3+ directly, then NaBH4 is added to measure total arsenic; As5+ concentration is determined by subtraction.

The research described relies on a high-quality, low-background-current substrate (BDD) modified with platinum microparticles for trace heavy-metal electroanalysis. Reproducing and extending this work requires precise control over diamond doping levels, surface termination, and metallic interface engineering—all areas where 6CCVD delivers engineering-grade solutions.

6CCVD MaterialRelevance to This Research
Heavy Boron-Doped PCD (BDD-grade)Direct substitute for the BDD substrate used here. Heavy doping (typically >1020 atoms/cm3) provides the metallic-like conductivity and low background current essential for ASV. Available in polycrystalline wafers up to 125 mm.
Optical Grade SCDAlternative substrate for single-crystal BDD variants, offering the lowest background current and widest potential window for ultra-trace detection (sub-ppb applications).
Custom Doping-Gradient BDDFor researchers seeking to optimize the boron concentration beyond standard levels, 6CCVD can tailor doping profiles during MPCVD growth.
  • Substrate Geometry: The paper used small BDD coupons. 6CCVD supplies plates and wafers up to 125 mm (PCD) and substrates up to 10 mm thick, enabling both single-electrode prototyping and multi-electrode array fabrication.
  • Metalization & Electrode Patterning: 6CCVD offers in-house metalization with Au, Pt, Pd, Ti, W, and Cu. For Pt-modified BDD electrodes, we can pre-deposit patterned Pt contact pads or seed layers via sputtering/evaporation, eliminating the wet-seeding step variability and improving reproducibility.
  • Laser Cutting & Dicing: Custom electrode geometries (disk, rectangular, interdigitated) can be precision-cut to specification, supporting both laboratory research and commercial sensor prototyping.
  • Surface Polishing: PCD surfaces can be supplied at Ra < 5 nm (inch-size), providing the smooth, uniform topography required for homogeneous Pt electrodeposition and reproducible ASV response.
  • 6CCVD’s in-house PhD team can assist with material selection, boron-doping concentration optimization, and metalization stack design for electrochemical sensor projects—including arsenic, heavy-metal, and biosensing platforms.
  • For applications requiring the combination of BDD + noble-metal nanostructures (Pt, Au, Pd), we can advise on seeding/electrodeposition protocols, adhesion layer selection (e.g., Ti underlayer for Pt), and calcination-compatible diamond grades.
  • Whether the goal is replicating the speciation analysis demonstrated in this paper (As3+/As5+ differentiation) or scaling it into a deployable sensor array, our team can recommend the optimal diamond grade, thickness (0.1 µm-500 µm), and metallization scheme.

For custom specifications, material consultation, or quotes on BDD substrates, metalized electrodes, or laser-patterned diamond sensor prototypes, visit 6ccvd.com or contact our engineering team directly.

View Original Abstract

Abstract High homogeneity and stability of platinum particles modified at boron-doped diamond (BDD) electrode was successfully performed using chemical seeding, followed by electrodeposition technique. The platinum particles on the surface of BDD characterized using Scanning Electron Microscopy-Energy Dispersive X-Ray Spectroscopy (SEM-EDS) showed that the coverage of Pt particles was 1.54 % on the BDD surface. These modified electrodes were used for As 3+ and As 5+ sensors using the anodic stripping voltammetry technique in 0.1 M phosphate buffer solution pH 6. Prior to the detection, pretreatment was required by the addition of 0.1 M NaBH 4 to change As 5+ to As 3+ . The optimum parameters were at a deposition potential of -500 mV, a deposition time of 150 s, and a scan rate of 200 mV/s. The voltammograms were linear for As 3+ and As 5+ (R 2 = 0.9797 and 0.9903, respectively) in the concentration range of 0 to 100 ppb. Estimated detection limits of 16.50 and 8.19 ppb for AS 3+ and As 5+ were achieved, respectively, suggesting that the method is promising for the specification detection of As 3+ and As 5+ .