Rapid determination of single substitutional nitrogen Ns concentration in diamond from UV-Vis spectroscopy
At a Glance
Section titled “At a Glance”| Metadata | Details |
|---|---|
| Publication Date | 2022-08-08 |
| Journal | Applied Physics Letters |
| Authors | Tingpeng Luo, Lukas Lindner, Rémi Blinder, Marco Capelli, Julia Langer |
| Institutions | RMIT University, Fraunhofer Institute for Applied Solid State Physics |
| Citations | 16 |
| Analysis | Full AI Review Included |
Rapid Determination of Single Substitutional Nitrogen in Diamond via UV-Vis Spectroscopy
Section titled “Rapid Determination of Single Substitutional Nitrogen in Diamond via UV-Vis Spectroscopy”Executive Summary
Section titled “Executive Summary”- Problem Solved: Established a robust, rapid, and replicable fitting protocol to determine single substitutional nitrogen (N0s) concentration directly from UV-Visible absorption spectroscopy, eliminating the need for complex, labor-intensive Electron Paramagnetic Resonance (EPR) measurements.
- Key Innovation: The method uses an “El-offset” spectrum from electronic-grade diamond as a baseline rather than relying on complex HPHT reference spectra. It accommodates CVD diamonds with varying defect concentrations and overlapping spectral features.
- Calibrated Metric: Cross-validated against EPR to define a precise absorption cross-section: σ = 1.96 ± 0.15 cm-1·ppm-1 (common logarithm) and σe = 4.51 ± 0.35 cm-1·ppm-1 (natural logarithm).
- Sensitivity Range: Capable of detecting nitrogen concentrations as low as 0.01 ppm and up to 30-50 ppm using standard 300 µm plates, significantly outperforming traditional FTIR spectroscopy.
- Practical Workflow: Converts raw transmission data into absolute N0s density using the calibrated absorption cross-section, enabling setup-independent quantification across different labs.
- Application Impact: Directly benefits NV-center quantum sensing research (magnetometry, biosensing, NMR) by providing rapid feedback on material quality and decoherence source concentration.
- Methodology: Utilized six (100)-oriented CVD diamond samples with nitrogen-doping levels ranging from 3.2 to 19.3 ppm, fitted with three Gaussian functions (270, 360, 520 nm), a ramp function, and an electronic-grade offset.
Technical Specifications
Section titled “Technical Specifications”| Parameter | Value | Unit | Context |
|---|---|---|---|
| Absorption Cross-Section (σ) | 1.96 ± 0.15 | cm-1·ppm-1 | Common logarithm basis |
| Absorption Cross-Section (σe) | 4.51 ± 0.35 | cm-1·ppm-1 | Natural logarithm basis |
| Cross-Section (σ) | 1.11 ± 0.09 × 10-17 | cm2 | Converted absolute value |
| Target Absorption Band | 270 | nm | N0s signature peak |
| Secondary Band 1 | 360 | nm | Vacancy clusters |
| Secondary Band 2 | 520 | nm | NVH0 centers |
| EPR Microwave Frequency | 9.84 | GHz | Calibration reference |
| UV-Vis Spectral Range | 200 - 800 | nm | Measurement window |
| Sample Concentration Range | 0.01 - 50 | ppm | Detection limits |
| EPR Measurement Error | ±6 | % | Combined statistical/fitting error |
| Fitting Error (l270) | ±1 | % | UV-Vis extraction precision |
| Carbon Atomic Mass (mC) | 1.99 × 10-23 | g | Conversion constant |
| Diamond Density (ρdia) | 3.51 | g/cm3 | Conversion constant |
| Ramp Function Exponent | -3 | — | R · λ-3 baseline model |
| Fitting Boundary (b270) | 268 - 272 | nm | Gaussian center bounds |
| Fitting Boundary (c270) | 13 - 27 | nm | Gaussian RMS width bounds |
| Sample Thickness (typical) | 300 | µm | Reference plate geometry |
Key Methodologies
Section titled “Key Methodologies”- Sample Preparation:
- Six (100)-oriented CVD diamond plates grown with varying nitrogen-doping levels.
- Both sides polished to enable transmission-mode UV-Vis measurement.
- EPR Calibration:
- Pre-characterized using Bruker ELEXSYS E580 spectrometer with super-high-Q probehead (ER4122 SHQE).
- Microwave frequency set to 9.84 GHz at room temperature.
- Spin-counting feature in xEPR software used for absolute concentration determination.
- UV-Vis Acquisition:
- Transmission (T) measured from 200-800 nm using PerkinElmer Lambda 950 spectrometer.
- Absorption coefficient (A) calculated via Lambert-Beer law: A = −log10(T)/d.
- Spectral Fitting Protocol (Five Components):
- Three Gaussian functions (gj(λ) = aj · exp(−(λ − bj)2 / (2cj2))) for bands at 270, 360, and 520 nm.
- One ramp function (r(λ) = R · λ-3) to model the decreasing absorption continuum.
- One “El-offset” spectrum (e(λ)) measured from an undoped electronic-grade diamond acting as baseline.
- Boundary Constraints:
- Gaussian center b270: bounded 268-272 nm.
- Gaussian RMS width c270: bounded 13-27 nm.
- Data Processing:
- Non-linear least squares fit applied to extract the 270 nm peak height (l270).
- Cutoff at 650 nm applied for samples with strong high-wavelength features (e.g., H2/NVN− centers).
- Linear regression between l270 and EPR concentration yields absorption cross-section.
6CCVD Solutions & Capabilities
Section titled “6CCVD Solutions & Capabilities”-
Applicable Materials:
- Nitrogen-Doped CVD Single Crystal Diamond (SCD): This research specifically relies on (100)-oriented single-crystal substrates with controlled nitrogen incorporation. Our Optical Grade SCD and custom nitrogen-doped variants provide the precise doping control required for NV-center formation research and N0s density calibration.
- Electronic Grade SCD (Low Nitrogen): The fitting protocol requires an undoped reference spectrum (“El-offset”). Our ultra-low nitrogen SCD wafers serve as an ideal baseline reference, with N content well below 1 ppb ensuring a clean spectral offset.
- Polycrystalline CVD Diamond (PCD): For applications requiring spatially resolved N0s mapping across larger areas (up to 125mm wafers), our PCD plates provide uniform characterization platforms.
-
Customization Potential:
- Precision Polishing: The Lambert-Beer calculation requires both-side polished samples with known thickness (d). 6CCVD offers Ra < 1nm polishing for SCD and Ra < 5nm for inch-size PCD, ensuring accurate absorption coefficient extraction.
- Custom Thickness: Samples were tested at 300 µm, but the method extends to thinner or thicker geometries. We supply SCD and PCD in thicknesses from 0.1 µm to 500 µm, and substrates up to 10mm, enabling optimization for any concentration range.
- Crystal Orientation Control: The paper specifies (100)-oriented substrates. We provide custom-oriented SCD substrates tailored to your epitaxial growth requirements.
- Wafer Dimensions: For high-throughput characterization campaigns or spatially resolved UV-Vis mapping, we deliver plates and wafers up to 125mm diameter.
-
Engineering Support:
- Reproducing this calibration protocol requires well-characterized reference materials with verified N0s concentrations. 6CCVD’s in-house PhD engineering team can assist with material selection for nitrogen density benchmarking, NV-magnetometry sensor development, and quantum metrology platforms.
- Whether your project targets magnetometry, biosensing, NMR spectroscopy, or scanning probe microscopy, we provide consultation on optimal nitrogen-doping levels to balance NV-center yield against decoherence sources.
-
Call to Action:
- For custom specifications, reference-grade diamond materials, or material consultation for your NV-center research, visit 6ccvd.com or contact our engineering team directly.
View Original Abstract
Single substitutional nitrogen atoms Ns0 are the prerequisite to create nitrogen-vacancy (NV) centers in diamonds. They not only serve as the electron donors to create the desired NV− center and provide charge stability against photo-ionisation but also are the main source of decoherence. Therefore, precise and quick determination of Ns0 concentration is a key advantage to a multitude of NV-related research in terms of material improvement as well as applications. Here, we present a method to determine the Ns0 concentration based on absorption spectroscopy in the UV-Visible range and fitting the 270 nm absorption band. UV-Visible spectroscopy has experimental simplicity and widespread availability that bear advantages over established methods. It allows a rapid determination of Ns0 densities, even for large numbers of samples. Our method shows further advantages in determining low concentrations as well as the ability to measure locally, which is highly relevant for diamonds with largely varying Ns0 concentrations in a single crystal. A cross-check with electron paramagnetic resonance shows high reliability of our method and yields the absorption cross section of the 270 nm absorption band σ=1.96±0.15 cm−1 ppm−1 (in common logarithm) or σe=4.51±0.35 cm−1 ppm−1 (in natural logarithm), which serves as a reference to determine Ns0 concentrations and makes our method applicable for others without the need for a known Ns0-reference sample and calibration. We provide a rapid, practical, and replicable pathway that is independent of the machine used and can be widely implemented as a standard characterization method for the determination of Ns0 concentrations.
Tech Support
Section titled “Tech Support”Original Source
Section titled “Original Source”References
Section titled “References”- 2020 - Nitrogen in diamond [Crossref]
- 2011 - Diamond photonics [Crossref]
- 2013 - The nitrogen-vacancy colour centre in diamond [Crossref]
- 2008 - Scanning magnetic field microscope with a diamond single-spin sensor [Crossref]
- 2009 - Diamonds with a high density of nitrogen-vacancy centers for magnetometry applications [Crossref]
- 2014 - Magnetometry with nitrogen-vacancy defects in diamond [Crossref]
- 2016 - Laser threshold magnetometry [Crossref]
- 2022 - Magnetic-field-dependent stimulated emission from nitrogen-vacancy centers in diamond [Crossref]
- 2014 - Nitrogen-vacancy centers in diamond: Nanoscale sensors for physics and biology [Crossref]
- 2016 - Diamond quantum devices in biology [Crossref]