A Compact Incubation Platform for Long‐Term Cultivation of Biological Samples for Nitrogen‐Vacancy Center Widefield Microscopy

FieldValue
JournalAdvanced Materials Technologies
Published2026-03-25
AuthorsAndre Pointner, Daniela Thalheim, Sarah Belasi, Lukas Heinen, Lucas R. Carnell
DOI10.1002/admt.70949

A compact incubation chamber supported HT29 human colorectal cancer cell viability and proliferation for 90 hours. It also enabled magnetic field imaging of labeled cells after 48 hours on a functionalized diamond surface. The 90-hour test used a cover slide, while the direct diamond test lasted 48 hours. The stage-top design preserves the inverted total internal reflection geometry of nitrogen-vacancy center widefield magnetometry while controlling temperature, carbon dioxide, and humidity.

The microscope used a [100] bulk diamond sample with a 1 μm thick, 1 ppm NV layer, fabricated by Quantum Diamonds. The sample was mounted on a polished glass cube. Total internal reflection directed 532 nm excitation into the diamond while reflecting the beam away from the biological sample and culture medium. The coupling angle was 36° to 39°, depending on the selected field of view.

Brightfield imaging used a 652 μm × 652 μm field of view. NV emission was evaluated over 318.5 μm × 318.5 μm at a laser intensity of 1 kW/cm². The camera signal was binned 2 × 2 to improve the signal-to-noise ratio. A dual-tone microwave scheme used a 1.5 MHz intermediate frequency and drove both nitrogen-15 hyperfine transitions. Interleaved images with and without microwave driving produced optically detected magnetic resonance contrast. The center of three resonances provided the local field along the sensing axis through the electron Zeeman effect. Rolling-ball subtraction removed low-frequency spatial noise and exposed dipole patterns from superparamagnetic iron oxide nanoparticles.

Incubation chamber design and environmental control

Section titled “Incubation chamber design and environmental control”

The chamber joined an acrylic cylinder to the glass cube with NOA61 optical adhesive. A rubber gasket and a lid sealed the culture volume. The authors chose a hybrid construction that combines SLA-printed BioMed White Resin parts with reusable elements. Ports admitted a controlled gas mixture and a water reservoir. Grooves at the top of the culture volume allowed limited gas exchange without opening the chamber to unrestricted evaporation.

A gas mixer supplied 5% CO₂ and 95% air. A temperature-controlled humidifier conditioned the mixture before it entered the chamber. The reservoir maintained oversaturated humidity and reduced medium evaporation. Condensate collected in an internal overflow and replenished the reservoir. The authors described condensation in the supply line as unavoidable, although heating the line to 37°C could reduce it.

A microcontroller used PID control of thick-film resistive heaters and NTC thermistors. Its process value was held near 34°C, slightly below 37°C. The lid was kept 2°C to 3°C above the incubation temperature to limit condensation. Calibration with deionized water found a stable 3°C gradient between the aluminum heater and the diamond surface. Sensors monitored the heaters and incubation atmosphere. This calibration allowed indirect control without placing a temperature probe at the diamond, where it could interfere with ODMR. The applied microwave power caused no measurable heating in either deionized water or culture medium. Magnets, stages, and optics remained outside the humid culture atmosphere.

The chamber floor was sealed with a microscope cover slide functionalized with fibronectin. A single suspension containing approximately 40,000 HT29 cells in 500 μL of medium filled the culture volume. Periodic brightfield images showed cell adhesion, morphology, and proliferation over 90 hours. The authors used these observations to confirm viability under constant incubation conditions.

Cells also showed difficulty adhering to bare diamond. Fibronectin and poly-L-lysine each improved adhesion in 48-hour tests. Poly-L-lysine was selected for later experiments because its electrostatic mechanism produced faster initial adhesion. The conclusion reports that fibronectin supported higher proliferation rates and provided a more natural extracellular matrix. Subsequent direct-on-diamond experiments used poly-L-lysine and magnetically labeled cells with superparamagnetic iron oxide nanoparticles following the authors’ earlier protocol.

At the end of the 48-hour diamond incubation, the magnetic field map showed clear dipole patterns from labeled cells. Two samarium cobalt ring magnets, separated by 70 mm, supplied a homogeneous static field of up to 30 mT along the selected NV axis. This field magnetized the nanoparticles along the selected axis.

Cell morphology changed during incubation, and proliferation altered the distribution of magnetic labels on the cell surface. The resulting field patterns showed corresponding deformations. The upper and lower image edges were noisy because excitation intensity decreased toward the field-of-view edges. Projection of the Gaussian beam at the total internal reflection angle produced an elliptical magnetic field of view.

Commercial stage-top and whole-microscope incubators were designed for conventional inverted microscopes with objective-based imaging. The authors state that this geometry is incompatible with the total internal reflection arrangement used for NV widefield magnetometry. The resulting lack of environmental control limited long-term studies of cellular dynamics. Their chamber places the culture volume at the sensing surface while keeping magnets, stages, and optics outside it.

The paper presents NV magnetic imaging as an alternative that avoids the photobleaching and phototoxicity that constrain long-term fluorescence imaging. The authors describe NV centers as offering unlimited photostability. The paper does not report a head-to-head comparison with a commercial incubator or a numerical gain in cell viability. Its reference point is the geometric incompatibility of existing incubators and the prior lack of environmental control during NV widefield measurements.

Engineering implications for diamond selection

Section titled “Engineering implications for diamond selection”

The demonstrated sensor couples several requirements that a replacement diamond must meet. It uses [100] orientation, a 1 μm deep, 1 ppm NV layer, and optical coupling to a glass cube. The sensing surface also carries cells and culture medium during microwave excitation. Bare diamond did not provide adequate adhesion, so surface functionalization is part of the operating design rather than an optional preparation step.

Temperature calibration must also account for the sensing plane. The measured 3°C heater-to-diamond gradient shows why a heater setpoint alone does not define the temperature at the cells. The chamber avoids an ODMR-interfering probe at the diamond. Any replacement geometry should preserve this separation and validate thermal behavior, gas exchange, and magnetic measurement quality.

The longest cultivation test lasted 90 hours. Direct cultivation on diamond was demonstrated for 48 hours, so the paper does not establish the same duration on the actual sensing surface. The text reports cell viability through morphology, adhesion, and proliferation observations. It does not report a numerical viability percentage, a time series of carbon dioxide concentration, or a numerical humidity value. The reported CO₂ control used a fixed 5% mixture and a water reservoir, with no stated closed-loop feedback for either gas.

The authors identify real-time CO₂ monitoring as a future improvement. They also propose perfusion for periods beyond 90 hours, with nutrient supply and waste removal, and multiwell chambers for parallel samples. Several constraints remain. Condensation in the gas supply line cannot yet be eliminated. Excitation decay makes the field-of-view edges noisy. Morphology-driven changes reduce labeling homogeneity. Data are available from the corresponding author upon reasonable request.

The paper’s stage-top NV magnetometer needs a [100] single-crystal diamond sensing plate with a 1 μm, 1 ppm NV layer, optically coupled to a polished glass cube. 6CCVD’s diamond materials catalog lists MPCVD-grown single-crystal plates up to 125 mm and polishing to Ra below 1 nm for single-crystal material. Because the paper does not state how its NV layer was produced, engineering support should confirm whether an MPCVD plate can meet the [100] orientation and 1 μm, 1 ppm NV-layer specification before an equivalent sensing plate is selected.

View Original Abstract

ABSTRACT Nitrogen‐vacancy (NV) centers in diamond provide a versatile quantum sensing platform for biological imaging through magnetic field detection, offering unlimited photostability and the ability to perform long‐term observations without photobleaching or phototoxicity. However, conventional stage‐top incubators are incompatible with the unique requirements for NV widefield magnetometry to study cellular dynamics. Here, we present a purpose‐built compact incubation platform that maintains precise environmental control of temperature, atmosphere, and humidity while accommodating the complex constraints of NV widefield microscopy. The system employs a 3D‐printed biocompatible chamber with integrated heating elements, temperature control, and humidified gas flow to create a stable physiological environment directly on the diamond sensing surface. We demonstrate sustained viability and proliferation of HT29 colorectal cancer cells over 90 hours of continuous incubation, with successful magnetic field imaging of immunomagnetically labeled cells after extended cultivation periods. This incubation platform enables long‐term cultivation and real‐time monitoring of biological samples on NV widefield magnetometry platforms, opening new possibilities for studying dynamic cellular processes using quantum sensing technologies.