Responsivity Study of Diamond X-ray Monitors with nUNCD Contact: Technical Documentation

Section titled “Responsivity Study of Diamond X-ray Monitors with nUNCD Contact: Technical Documentation”
  • Application: X-ray beam position monitors for synchrotron beamlines (0.2-28 keV photon range), developed at Brookhaven National Laboratory.
  • Device Architecture: Symmetric diamond detector utilizing nitrogen-doped ultrananocrystalline diamond (nUNCD) as electrodes on both sides of a CVD single-crystal diamond substrate, replacing conventional platinum (Pt) contacts.
  • Key Achievement: nUNCD contact demonstrated responsivity comparable to theoretical predictions (within 4% deviation), with successful charge collection at a lower bias of 50V versus the 100V required for Pt-contact devices.
  • Performance Advantage: At the carbon K-edge (~284 eV), the nUNCD contact outperformed the Pt-contact detector due to significantly lower x-ray absorption (Pt M-edges cause spectral artifacts and signal loss).
  • Substrate Specs: Electronic grade SCD (~ppb nitrogen), 4 mm × 4 mm × 0.3 mm, with 200-500 nm nUNCD layers grown via MPCVD.
  • Characterization: Validated at three NSLS beamlines (U3C, X8A, X15A) under varying vacuum conditions and biasing modes (DC and square-wave duty cycles for electron detrapping).
ParameterValueUnitContext
Substrate MaterialElectronic Grade SCD—[N] ~ppb level purity
Substrate Dimensions4 × 4 × 0.3mmStandard BNL test geometry
nUNCD Layer Thickness200 - 500nmGrown via MPCVD at ANL
Fitted nUNCD Thickness480nmIncluding dead carbon layer effect
Fitted Diamond Bulk Thickness380µmExtracted from 4-6 keV tail fitting
Tested Photon Energy Range0.2 - 28keVMulti-beamline validation
Bias Voltage (nUNCD)50VFor >95% charge collection
Bias Voltage (Pt reference)100VRequired for full carrier extraction
Mean Ionization Energy (W)13.6eV4% deviation from literature 13.3 eV
X-ray Attenuation Length (350 eV)0.0878µmDiamond bulk, CXRO data
X-ray Attenuation Length (1 keV)1.2888µmDiamond bulk, CXRO data
X-ray Attenuation Length (19 keV)7129.76µmDiamond bulk, CXRO data
Minimum Transmission Energy1350eVFor full traversal of 300 µm diamond
Beam Intensity Range10-9 - 10-6W/mm2Incident beam calibration range
U3C Vacuum Level10-7torrLow-energy soft x-ray beamline
X8A Vacuum Level10-6torrMid-range monochromatic beamline
Detector Biasing ModesDC / Square Wave—With adjustable magnitude, frequency, and duty cycle
  1. Substrate Preparation: Electronic grade CVD single-crystal diamond (4 mm × 4 mm × 0.3 mm, ~ppb nitrogen) selected as bulk detector material.
  2. nUNCD Layer Deposition: Nitrogen-doped ultrananocrystalline diamond layers (200-500 nm) grown on both faces via MPCVD at Argonne National Laboratory using Ar/CH4/N2 plasma chemistry to achieve semi-metallic conductivity.
  3. Pre-Fabrication Characterization: X-ray white-beam topography at NSLS Beamline X19C and polarized-light birefringence imaging to verify crystal quality and identify slip bands.
  4. Multi-Beamline Validation:
    • U3C: Monochromatic beam, 0.2-1 keV, 10-7 torr vacuum.
    • X8A: Monochromatic beam, 1-6.5 keV, 10-6 torr vacuum.
    • X15A: 6.5-28 keV, in-air configuration.
  5. Incident Power Calibration: Silicon photodiodes (25 µm at U3C, 52 µm at X8A); air-filled ionization chamber at X15A.
  6. Electrical Readout: DC and pulsed square-wave biases with tunable duty cycles applied for hole/electron collection and detrapping.
  7. Theoretical Modeling: Modified Beer-Lambert responsivity equation incorporating nUNCD filter thickness, dead carbon layer, and bulk diamond attenuation.

Applicable Materials:

  • Electronic Grade SCD (Optical Grade): The electronic grade single-crystal diamond substrate used here (~ppb nitrogen) is directly within 6CCVD’s MPCVD production scope, with thicknesses from 0.1 µm to 500 µm and dimensions scalable to 125 mm for polycrystalline formats.
  • Ultrananocrystalline Diamond (UNCD) / nUNCD Layers: 6CCVD’s MPCVD platform can deposit nanocrystalline and ultrananocrystalline diamond films with tunable sp2/sp3 ratios. Nitrogen doping for semi-metallic conductivity is achievable through Ar/CH4/N2 plasma recipes.
  • Boron-Doped Diamond (BDD): For alternative conductive contact schemes or bulk-conducting detector designs, heavy boron-doped polycrystalline diamond is available in thicknesses up to 500 µm.

Customization Potential:

  • Tailored Thickness Control: The paper notes that nUNCD thickness variation (200-500 nm) creates non-uniform response maps at low energies (halo artifacts). 6CCVD offers precision thickness control and in-house metrology for sub-µm layer uniformity critical to detector-grade responsivity.
  • Custom Metalization: For hybrid contact architectures (e.g., nUNCD + bonding pads), 6CCVD provides internal Au, Pt, Pd, Ti, W, and Cu metallization with customizable pad geometries and laser-cut form factors.
  • Laser Cutting & Dicing: Custom shaping of 4 mm × 4 mm plates or wafer-scale substrates into detector geometries with tight dimensional tolerances.

Engineering Support:

  • 6CCVD’s in-house PhD engineering team can collaborate on detector-grade diamond selection, including SCD purity grading, nUNCD conductivity optimization, and electrode interface engineering for synchrotron, medical imaging, and beam diagnostic applications.
  • Available consultation on dead carbon layer mitigation, MPCVD growth parameter optimization, and bias-dependent charge collection modeling.

Call to Action:

For custom specifications, metallization stack design, or material consultation for your x-ray detector or beam diagnostics project, visit 6ccvd.com or contact our engineering team directly.

View Original Abstract

Nitrogen doped ultrananocrystalline diamond (nUNCD) grown on the surface of a CVD single crystal diamond is tested at various beamlines covering an x-ray photon energy range of 200eV to 28 keV. The nUNCD has much lower x-ray absorption than metal contacts and is designed to improve the performance of our device. The responsivity of nUNCD diamond x-ray detector is compared with the conventional platinum coated diamond x-ray beam position monitor and the results are presented in this paper.