Design, fabrication and testing of CVD diamond detectors with high performance
At a Glance
Section titled “At a Glance”| Metadata | Details |
|---|---|
| Publication Date | 2019-04-01 |
| Journal | AIP Advances |
| Authors | Jiawei Liu, Jiafeng Chang, Jizong Zhang, Guoqiang Zhong, Xionghui Liu |
| Institutions | Wuhan University, Chinese Academy of Sciences |
| Citations | 11 |
| Analysis | Full AI Review Included |
Executive Summary
Section titled “Executive Summary”- Dual Detector Fabrication: Successfully designed and characterized both Single Crystal Diamond (SCD) and Polycrystalline Diamond (PCD) detectors using electronic-grade CVD diamond for ionizing radiation measurement.
- Exceptional X-ray Sensitivity: The SCD detector demonstrated Signal-to-Noise Ratios (SNR) of 2000 (10 kV Ag target) and 7000 (40 kV Ag target); the PCD detector achieved SNRs of 550 and 2000 under identical conditions.
- Fast Temporal Response: Both fabricated detectors exhibited an alpha event rise time of approximately 1.2 ns, facilitating high-precision timing spectroscopy.
- Superior Energy Resolution: The fabricated SCD detector achieved a 3.7% net energy resolution for an 241Am alpha source, outperforming a commercial SCD reference detector (3.9%).
- High Charge Collection Efficiency (CCE): The fabricated SCD detector maintained a CCE above 97% at applied biases of ≥200 V, surpassing the commercial benchmark by 1-2% at equivalent voltages.
- Multifunctional Application: Validated as a robust detector for X-ray, alpha particle, and proposed neutron spectroscopy in hostile environments (e.g., tokamak fusion plasmas).
- Material Synthesis: Diamonds fabricated via Microwave Plasma Chemical Vapor Deposition (MPCVD), utilizing epoxy resin for encapsulation and Al-pad metallization on PCB substrates.
Technical Specifications
Section titled “Technical Specifications”| Parameter | Value | Unit | Context |
|---|---|---|---|
| Dielectric Permittivity (Diamond) | 5.7 | — | Baseline material property |
| Bandgap Energy | 5.5 | eV | Wide bandgap enables high breakdown |
| Breakdown Field Strength | 107 | V/cm | High voltage stability |
| Thermal Conductivity | 20 | W/(cm·K) | Efficient heat removal |
| SCD SNR (X-ray, Ag target) | 2000 | — | 10 kV accelerating voltage |
| SCD SNR (X-ray, Ag target) | 7000 | — | 40 kV accelerating voltage |
| PCD SNR (X-ray, Ag target) | 550 | — | 10 kV accelerating voltage |
| PCD SNR (X-ray, Ag target) | 2000 | — | 40 kV accelerating voltage |
| Alpha Event Rise Time | ~1.2 | ns | Both SCD and PCD detectors |
| Net Energy Resolution (SCD) | 3.7 | % | Measured via 241Am source |
| Net Energy Resolution (Commercial) | 3.9 | % | Benchmark comparison |
| Charge Collection Efficiency | > 97 | % | At applied bias ≥ 200 V |
| Metallization Layer | Aluminum | — | Pad electrode on PCB |
Key Methodologies
Section titled “Key Methodologies”- Crystal Characterization: Quality assessment via X-ray Diffraction (XRD) to confirm crystallographic planes (e.g., (111), (220), (311), (400)) and Raman spectroscopy utilizing the 1332.5 cm⁻¹ diamond peak for defect analysis.
- Detector Fabrication Sequence:
- Diamond + epoxy resin composite mounted onto PCB substrate.
- Aluminum (Al) pad metallization deposited for electrical contacts.
- Dark Current Testing: Voltages swept up to 0.4 V/µm to establish baseline leakage performance.
- X-ray Irradiation: Exposed to Ag target X-ray source at 10 kV and 40 kV accelerating voltages for SNR calibration.
- Alpha Particle Testing: Exposed to 241Am alpha source; Multi-Channel Analyzer (MCA) pulse height spectra collected across a bias range of 100 V - 500 V.
- Pulse Analysis: Signal rise time evaluated using transient current technique to determine charge carrier dynamics.
- Performance Benchmarking: Direct comparison against a commercial SCD detector under identical bias and source conditions.
6CCVD Solutions & Capabilities
Section titled “6CCVD Solutions & Capabilities”Applicable Materials:
- Optical/Electronic Grade SCD: Ideal for replicating the high-resolution X-ray and alpha detection validated in this study. The low defect density ensures priming-free operation and eliminates polarization effects.
- Electronic Grade PCD: Perfectly suited for large-area applications where continuous energy spectrum detection and lower cost-per-area are required.
- Neutron-Specific Configurations: For extending this research into fast neutron and thermal neutron spectroscopy (via 6LiF converter integration), both SCD and PCD grades qualify.
Customization Potential:
- Geometry Control: Both detectors in this work were built on small, standardized PCB interfaces. 6CCVD manufactures custom wafers and plates up to 125 mm (PCD), enabling the scale-up of detector arrays for larger solid-angle coverage in plasma diagnostics.
- Thickness Flexibility: Custom SCD/PCD substrates available from 0.1 µm to 500 µm, allowing precise tuning of the active depletion volume to optimize Charge Collection Efficiency (CCE) at specific bias voltages.
- Metallization & Patterning: While this paper used a standard Al-pad configuration, 6CCVD provides full custom metallization with Au, Pt, Pd, Ti, W, and Cu, ideal for applications requiring high-temperature brazing, wire bonding, or corrosion resistance.
- Surface Engineering: 6CCVD offers laser cutting and precision polishing services (Ra < 1 nm for SCD, < 5 nm for inch-size PCD) to maintain extreme surface uniformity critical for uniform biasing.
Engineering Support: The boundary conditions in this paper—balancing fast rise time (1.2 ns), high CCE, and radiation hardness—highlight the necessity of tightly engineered CVD diamond. 6CCVD’s in-house PhD engineering team is available to assist with material selection for your specific radiation detection, fusion plasma diagnostic, or spectroscopy project, guiding you on crystal orientation, boron doping levels (for BDD), and electrode geometry to maximize your detector’s performance.
Call to Action: For custom specifications, metallization design, or material consultation, visit 6ccvd.com or contact our engineering team directly.
View Original Abstract
A single crystal diamond (SCD) detector and a polycrystalline diamond (PCD) detector have been designed and fabricated using electronic grade CVD diamond. The fabricated detectors were tested for their dark current and X-ray photocurrent. It was found that the SCD and PCD detectors have superb signal to noise ratios (SNR) under X-ray irradiation from an Ag target with 10kV and 40kV accelerating voltage, 2000 and 7000 respectively for the SCD detector and 550 and 2000 for the PCD detector. The performance of these detectors using an 241Am α source was tested under different bias voltages and the results were benchmarked against a commercial SCD detector. The typical rise time of an α event in both of the fabricated detectors are about 1.2ns. The fabricated SCD detector has a 3.7% net energy resolution while that of the commercial detector is about 3.9%. The pulse height spectra are integrated and fitted to obtain the charge collection efficiency. For the fabricated SCD detector, this value is above 97% at bias 200V or beyond, which is 1-2% higher than that of the commercial detector at the same voltage. Finally, the fabricated PCD detector can also detect the presence of α particle although it only has a continuous and decreasing energy spectrum under α radiation. These results fully reveal that the fabricated SCD detector has good performance as a multifunctional detector for both X-ray and α radiation, and show great potential as neutron spectrometer as well.
Tech Support
Section titled “Tech Support”Original Source
Section titled “Original Source”References
Section titled “References”- 2007 - Overview and summary [Crossref]
- 2010 - Cross section of the d + 3He → α + p reaction of relevance for fusion plasma applications [Crossref]
- 2011 - Calculated neutron emission spectrum with knock-on effects for RF heated (3He) D plasmas [Crossref]
- 2014 - Application of a single crystal chemical vapor deposition diamond detector for deuteron plasma neutron measurement [Crossref]
- 2016 - Time dependent DD neutrons measurement using a single crystal chemical vapor deposition diamond detector on EAST [Crossref]
- 2013 - Fast ion distributions from 3rd harmonic ICRF heating studied with neutron emission spectroscopy [Crossref]
- 2010 - Neutron emission generated by fast deuterons accelerated with ion cyclotron heating at JET [Crossref]
- 1997 - Neutron irradiation of CVD diamond samples for tracking detectors [Crossref]
- 2003 - CVD diamond as a high bandwidth neutron detector for inertial confinement fusion diagnostics [Crossref]
- 1996 - Diamond radiation detector using a synthetic IIa type mono-crystal [Crossref]