Performance of the Lancelot Beam Position Monitor at the Diamond Light Source
At a Glance
Section titled āAt a Glanceā| Metadata | Details |
|---|---|
| Publication Date | 2017-12-21 |
| Journal | Journal of Instrumentation |
| Authors | H. Chagani, Tomas B. Garcia-Nathan, Chongyun Jiang, Anton Kachatkou, J. Marchal |
| Institutions | Diamond Light Source, University of Manchester |
| Citations | 1 |
Abstract
Section titled āAbstractāThe Lancelot beam position and profile monitor records the scattered radiation off a thin, low-density foil, which passes through a pinhole perpendicular to the path of the beam and is detected by a Medipix3RX sensor. This arrangement does not expose the detector to the direct beam at synchrotrons and results in a negligible drop in flux downstream of the module. It allows for magnified images of the beam to be acquired in real time with high signal-to-noise ratios, enabling measurements of tiny displacements in the position of the centroid of approximately 1 μm. It also provides a means for independently measuring the photon energy of the incident monoenergetic photon beam. A constant frame rate of up to 245 Hz is achieved. The results of measurements with two Lancelot detectors installed in different environments at the Diamond Light Source are presented and their performance is discussed.